Endocardiális myocardiális pacemaker elektródák hatása a szívizomzatra
Title in English
The influence of myocardial pacemaker elektrodes to the heart mosculatre
Panel
Plant and animal breeding
Department or equivalent
Belgyógyászati Tanszék és Klinika (Hungarian University of Agriculture and Life Sciences)
Participants
Balogh, Emese Dudás, Gábor Petneházy, Örs Sótonyi, Péter
Starting date
2002-01-01
Closing date
2005-12-31
Funding (in million HUF)
6.694
FTE (full time equivalent)
0.00
state
closed project
Final report
Results in Hungarian
Huszonkét 3-4 hónapos, sertésbe összesen 28 pacemaker elektródát ültettünk be. A kísérleti állatokat 3 hónapon keresztül tartottuk. A kísérlet során az állatok több mint felében tapasztaltunk súlyos fokú komplikáció kialakulását, így végül 10 állatba beültetett 13 elektróda elektromos adatait tudtuk feldolgozni. Nem találtunk szignifikáns különbséget sem a kétféle fixációjú ingerelt elektróda jelnagysága, sem küszöb ingerlési potenciálja között. Nem volt szignifikáns különbség az aktív fixációjú endocardiális ingerelt és ingerlés nélküli elektródák ellenállásában, küszöb ingerlési potenciáljában, ill. jelnagyságában sem. Szöveti reakciókat 20 sertésből származó szívizom mintákon vizsgáltuk. A vizsgált elváltozások előfordulásában sem a pacemaker elektróda típusa, sem az ingerlő feszültség nem volt meghatározó. A nem ingerelt elektródák végénél és környezetükben sokszor az ingereltekével azonos reakciók jelentek meg. Így azt a következtetést vontuk le, hogy a pacemaker elektródák körül kialakuló reakcióban az elektródavégek elektromos tulajdonságainak nincs meghatározó szerepe. Mostani vizsgálatainkban nem tudtunk a bekövetkező szövettani reakció mértéke és az elektróda elektromos paramétereinek megváltozása között sem összefüggést találni. Kutatásunk során vizsgáltuk a setések más cardiovascularis paramétereit, továbbá a cardialis memória előfordulását is. Leírtuk a pacemaker beültetések komplikációit és ezek megelőzésének lehetőségeit is.
Results in English
Pacemaker implantations were performed in twenty-two 3-4 months old swine. The animals were kept for 3 months. Severe complications were met in more than half of the experimental animals, thus 13 electrodes from 10 animals were ready for electronic testing at the end of the study. We didn’t find significant difference between the threshold potential and the mean signal amplitude of the active (screw in) and passive fixated endomyocardial electrodes. Similarly, no significant differences were found between the threshold potential, mean signal amplitude and impedance of the paced and non-paced screw in electrodes. Histological samples were obtained from twenty pacemaker implanted swine. There was no association between the observed histological lesions and the electrode types or the pacing voltages. Similar histological lesions were found around the tip of paced and non-paced screw-in endomyocardial electrodes. Therefore we suggest that the histological lesions are determined almost exclusively by the electrode material and not by the fixation type or pacing parameters. We couldn’t find association between the different histological lesions and the measured electrode parameters during this study. We also investigated and described other cardiovascular parameters of the pig and the occurrence of cardiac memory during pacing. We analyzed the possible causes and prevention of pacemaker implantation complication in this species.
Manczur Ferenc, Hetyey Csaba, Reiczigel Jenő: Kutyák szívbetegségeinek előfordulása Magyarországon (1997-2000), Magyar Állatorvosok Lapja, 2003
Ferenc Manczur, Csaba Hetyey, Jenő Reiczigel: Prevalence of cardiological diseases in Hungary (1997-2000), Journal of Veterinary Internal Medicine, 2003
Events of the project
2013-01-30 15:00:22
Kutatóhely váltás
A kutatás helye megváltozott. Korábbi kutatóhely: Anatómiai és Szövettani Tanszék (Szent István Egyetem), Új kutatóhely: Belgyógyászati Tanszék és Klinika (Szent István Egyetem).